| LM-80 Tester |
| Conditions |
Tcase = 55℃, 85℃, third temp. selected by the manufacturer |
| Duration |
≧6000hrs |
| Test Standard |
IESNA LM-80 |
| Target Device |
High Power LED , Medium Power LED |
| Application |
LM-80 Test , Tcase precise temp. stress test. |
| Thermal bench performance |
Temperature range: 30℃~125 ℃
control stability:±0.5℃ (single LED / bench)
control stability:±2℃ (several LEDs / bench) |
| Environmental chamber performance |
Temperature range: 30℃~125 ℃
control stability: ±2℃
Humidity range ≦ 65% R.H. |
| Control Software |
window-based user interface |
| Control Interface |
RS485 / GPIB / Industrial IO card |
| Power |
3 phase, 220V,
25A / 50A (depend on ATC bench amount) |
| Notes |
◆ Case Temp. shall be controlled in range Tcase ≧ Tstress - 2℃
◆ Surrounding air should be maintained to within Ta ≧Tstress - 5℃ during testing
◆ Humidity ≦ 65% R.H.
◆ Airflow shall be minimized because of heat characteristics that differ due to variation in airflow.
◆ The Tcase、Ta、If should be monitored & recorded. Once the test conditions is not match, the test duration timer should stop accumulating. |